|
External standard (ES) calibration
advantage: very simple calibration method
disadvantage: high susceptibility to systematic errors
There are
more elaborate calibration methods, which are less susceptible to systematic errors, are required in environmental trace analysis:
- internal standard calibration
- calibration by standard addition
three most common sources of systematic errors:
(i) signal drift
temporal changes in the response of the detector to a constant analyte concentration
|
|
(ii) matrix effects
components of the sample matrix (e.g., dissolved organic matter) suppress the response of the detector;
this effect is not accounted for by calibration with external standards, which are prepared in pure solvents
|
|
(iii) analyte loss
loss of analyte during sample preparation is not accounted for by calibration with external standards,
unless the standards are also carried through the sample preparation steps
|
|
|